Suppr超能文献

关于治疗用X射线设备的输出变化

On output variations of therapeutic x-ray equipment.

作者信息

Zaránd P

出版信息

Strahlentherapie. 1979 Mar;155(3):200-3.

PMID:432892
Abstract

The author has measured the exposure rate of 56 orthovolt X-ray units. A comparison of consecutive measurements resulted in a 4.2% standard deviation which is mainly due to output variation of the X-ray equipment itself and due to the variation of mains supply. The repeatability of an irradiation controlled by a timer is therefore +/- 9% (on the 95% confidence level), which is much lower than recommended by the ICRU or stipulated in Hungarian National Standards. This accuracy is however sufficient in non-tumor therapy. The output variation among therapeutic X-ray units is high (sigma x approximately 22%) and correlates with the HVL measurements. The advantage of a central dosimetric service is the reduction of the number of unnecessary radiation qualities and superfluous combinations of radiation qualities, tube current and FSD.

摘要

作者测量了56台深部X射线治疗机的照射率。连续测量结果的标准差为4.2%,这主要归因于X射线设备本身的输出变化以及电源供应的变化。因此,由定时器控制的照射的重复性为±9%(在95%置信水平下),这远低于国际辐射单位与测量委员会(ICRU)的建议或匈牙利国家标准的规定。然而,这种精度在非肿瘤治疗中是足够的。治疗用X射线机之间的输出变化很大(标准差约为22%),并且与半值层(HVL)测量值相关。中央剂量测定服务的优势在于减少了不必要的辐射质数量以及辐射质、管电流和焦皮距(FSD)的多余组合。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验