Development of a reversible vas deferens occlusive device. VI. Long-term evaluation of flexible prosthetic devices.

作者信息

Brueschke E E, Zaneveld L J, Kaleckas R A, Wingfield J R

出版信息

Fertil Steril. 1979 May;31(5):575-86. doi: 10.1016/s0015-0282(16)44005-7.

Abstract

Fifty-three dogs received implants of several types of flexible devices containing valving mechanisms. These devices were constructed entirely of silicone rubber with the exception of the valve stem, which was made of stainless steel. Generally, the devices were (1) implanted in the open mode and left this way for 27 to 44 months, (2) implanted in the closed mode and kept this way for 11 to 12 months before the valves were reopened, (3) implanted in either the closed or open mode and cycled to the opposite mode every 3 months (four or five cycles), or (4) implanted in either the closed or open mode and cycled to the opposite mode every 6 months (two or three cycles). Different implant methods were also evaluated. Semen analyses were regularly performed on all of the dogs, and a number of the animals were bred during the experiments. It can be concluded that (1) the devices can be opened and closed successfully over long periods, respectively allowing and preventing sperm transport; (2) the breeding ability of the animals is not impaired while the devices are in the open mode, independent of the type of device; (3) the devices do not result in an enhanced incidence of congenital abnormalities in the offspring; and (4) the success rate of device performance does not depend on the method of implanatation used. Thus, the results clearly indicate that the basic mechanism of a soft, reversible valve is a feasible approach to conception control.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验