Kiehn T E, Brennan K, Ellner P D
Appl Microbiol. 1974 Oct;28(4):668-71. doi: 10.1128/am.28.4.668-671.1974.
Clinical isolates (869) and stock cultures (35) of Enterobacteriaceae were tested in parallel with the Minitek and conventional systems. The Minitek correctly identified 822 of 904 cultures. When a deoxyribonuclease plate was inoculated along with the Minitek, it was possible to speciate Enterobacteriaceae within 24 h. False-positive hydrogen sulfide reactions were the major fault with this system. Reactions were clear-cut and easy for technologists to read.