Finley P R, Hicks M J, Williams R J, Hinlicky J, Lichti D A
Clin Chem. 1979 Nov;25(11):1909-14.
We describe the measurement of rheumatoid factor in human sera with a rate nephelometer. The National Reference Preparation for Rheumatoid Factors is used to calibrate the assay in International Units. We used Hyland Positive Control, Level I, as a secondary standard. The standard curve is exponential, but is linear when plotted on log-log graph paper. Aggregated immune globulin (IgG) is the antigen used to detect rheumatoid factor (IgM-class antibody to IgG). The rate reaction measures the rate of increase in light-scatter by the antigen-antibody complexes; the reaction takes place in 17 to 20 s. Precision, linearity, and accuracy are excellent. Results agree well with those for a commonly used latex precipitation test. The advantages of speed, quantification in International Units, and superior discrimination of concentration as compared to serological titration provide a more reliable test for use in the diagnosis and treatment of rheumatoid arthritis.
我们描述了使用速率散射比浊法测定人血清中类风湿因子的方法。类风湿因子国家参考制剂用于以国际单位校准该检测。我们使用海兰一级阳性对照作为二级标准品。标准曲线呈指数关系,但在对数-对数坐标纸上绘制时呈线性。聚合免疫球蛋白(IgG)用作检测类风湿因子(针对IgG的IgM类抗体)的抗原。速率反应测量抗原-抗体复合物光散射增加的速率;反应在17至20秒内发生。精密度、线性和准确性都非常好。结果与常用的乳胶沉淀试验结果非常一致。与血清学滴定相比,该方法具有速度快、以国际单位定量以及对浓度有更好区分度的优点,为类风湿关节炎的诊断和治疗提供了更可靠的检测方法。