Gunnarsson B, Asaba H, Kiibus A, Söderborg B, Wiklund S, Bergström J
Clin Nephrol. 1979 Sep;12(3):117-21.
We have estimated the degree of fibrin deposition as well as changes in dialyzer efficiency of Gambro Optima 13.5 mu dialyzers, re-used up to six times in six regular dialysis patients. After each dialysis the dialyzer was rinsed and left filled with formaldehyde. The patients were given I-125-fibrinogen before the study. The radioactivity from the dialyzer was continuously measured during each dialysis and rinsing procedure. Urea and creatinine clearances were determined 30 min after start of each dialysis. After the last re-use the dialyzer membranes were removed and divided into sections and the radioactivity of each section was measured. Twice as much fibrin deposition was found near the blood inlet of the dialyzer as near the outlet. A higher fibrin deposition was found after each re-use than after the initial dialysis. The clearances of urea and creatinine were scarcely influenced by the re-use. The half-life of fibrinogen in the patients was in the same range as in normal subjects.
我们评估了Gambro Optima 13.5 mu透析器的纤维蛋白沉积程度以及透析效率变化,该透析器在6例常规透析患者中重复使用多达6次。每次透析后,透析器进行冲洗并充满甲醛。在研究前给患者注射I-125-纤维蛋白原。在每次透析和冲洗过程中持续测量透析器的放射性。每次透析开始30分钟后测定尿素和肌酐清除率。在最后一次重复使用后,取出透析器膜并分成几部分,测量每部分的放射性。发现透析器血液入口处附近的纤维蛋白沉积量是出口处附近的两倍。每次重复使用后的纤维蛋白沉积量都比初次透析后高。尿素和肌酐清除率几乎不受重复使用的影响。患者体内纤维蛋白原的半衰期与正常受试者处于相同范围。