经皮连续波多普勒计算机测定心输出量变化的准确性

Accuracy of determination of changes in cardiac output by transcutaneous continuous-wave Doppler computer.

作者信息

Rose J S, Nanna M, Rahimtoola S H, Elkayam U, McKay C, Chandraratna P A

出版信息

Am J Cardiol. 1984 Nov 1;54(8):1099-101. doi: 10.1016/s0002-9149(84)80151-4.

Abstract

The value of a previously validated portable, continuous-wave Doppler computer was assessed for measuring changes in cardiac output (CO). Simultaneous thermodilution and Doppler CO values were measured in triplicate in 16 patients undergoing clinical intervention with vasodilator therapy. A continuous-wave Doppler transducer was placed in the suprasternal notch and directed toward the ascending aorta and angled until the maximal velocity signal was obtained. The correlation coefficient was 0.92 (standard error of the estimate [SEE] = 0.48 liter/min) at rest; with intervention it was 0.88 (SEE = 0.52 liter/min). Our data indicate that the Doppler computer technique, when used in selected patients, is reliable in detecting changes in CO after vasodilator therapy. It may be of value in clinical situations in which hemodynamic monitoring is impractical.

摘要

对一台先前经验证的便携式连续波多普勒计算机在测量心输出量(CO)变化方面的价值进行了评估。对16例接受血管扩张剂治疗临床干预的患者,同时采用热稀释法和多普勒法测量CO值,各测量3次。将连续波多普勒换能器置于胸骨上切迹处,朝向升主动脉,调整角度直至获得最大速度信号。静息时相关系数为0.92(估计标准误差[SEE]=0.48升/分钟);干预后为0.88(SEE=0.52升/分钟)。我们的数据表明,多普勒计算机技术应用于特定患者时,在检测血管扩张剂治疗后CO的变化方面是可靠的。在血流动力学监测不实用的临床情况下可能具有价值。