Suppr超能文献

与低温恒温器切片相关的切片内和切片交叉处厚度变异性的干涉测量分析。

Interferometric analysis of intrasection and intersection thickness variability associated with cryostat microtomy.

作者信息

Anthony A, Colurso G J, Bocan T M, Doebler J A

出版信息

Histochem J. 1984 Jan;16(1):61-70. doi: 10.1007/BF01003436.

Abstract

Mach-Zehnder interferometric measurements were used to assess the extent of section thickness variability (inter- and intrasection) associated with cryostat microtomy of adrenal sections over a typical working range of 10-20 micron. Sections were obtained using a Bright's Cambridge rocking type and a Damon rotary type cryostat microtome to allow comparative analyses. The effective thickness of tissue sections after being mounted onto slides by flash drying was reduced by 90% relative to microtome section thickness setting. A linear relationship between measured thickness and microtome setting was obtained with both instruments. Thickness variability between replicate sections over the range of microtome settings approximated 11% for the rocking microtome and 5% with the rotary microtome. Average intrasection variability was found to be 7% for rocking microtome sections and 4% for sections obtained with the rotary microtome. However, this variability is a negligible source of error in cytophotometric analyses, providing replicate sections are used and an adequate number of measurements are made on mask-delimited individual cells or tissue specimen areas.

摘要

马赫-曾德尔干涉测量法用于评估在10-20微米的典型工作范围内,与肾上腺切片低温切片术相关的切片厚度变异性(切片间和切片内)程度。使用布莱特剑桥摇摆式和达蒙旋转式低温切片机获取切片,以便进行比较分析。通过快速干燥将组织切片安装到载玻片上后,其有效厚度相对于切片机切片厚度设置减少了90%。两种仪器均获得了测量厚度与切片机设置之间的线性关系。在切片机设置范围内,重复切片之间的厚度变异性对于摇摆式切片机约为11%,对于旋转式切片机为5%。发现摇摆式切片机切片的平均切片内变异性为7%,旋转式切片机获得的切片为4%。然而,在细胞光度分析中,这种变异性是一个可忽略不计的误差来源,前提是使用重复切片并对经掩膜界定的单个细胞或组织标本区域进行足够数量的测量。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验