Mehta S
Ann R Coll Surg Engl. 1984 Nov;66(6):426-7.
Although the incidence of cuff induced tracheal damage has been considerably reduced by judicious use of tracheal tubes with large-volume, low-pressure cuffs, aspiration continues to be a major problem. A study was conducted to determine the maximum hydrostatic pressure that can be produced by a column of liquid above the tracheal cuff. The vertical and horizontal distance between the upper central incisor teeth and suprasternal notch was measured in 200 supine adult subjects. The maximum vertical and horizontal distance between the upper central incisor teeth and suprasternal notch was 9 cm and 21 cm respectively. The implication of these findings are discussed and it is suggested that a fixed intracuff pressure in the range of 2.5 to 3kPa should be used to obtain an optimal tracheal seal.
尽管通过明智地使用大容量、低压套囊的气管导管,套囊引起的气管损伤发生率已大幅降低,但误吸仍然是一个主要问题。进行了一项研究以确定气管套囊上方的液柱所能产生的最大静水压力。在200名仰卧位成年受试者中测量了上中切牙与胸骨上切迹之间的垂直和水平距离。上中切牙与胸骨上切迹之间的最大垂直和水平距离分别为9厘米和21厘米。讨论了这些发现的意义,并建议使用2.5至3kPa范围内的固定套囊内压力以获得最佳的气管密封。