Suppr超能文献

脉冲多普勒超声心动图检测主动脉瓣反流和二尖瓣反流的敏感性与特异性。

Sensitivity and specificity of pulsed Doppler echocardiography in detection of aortic and mitral regurgitation.

作者信息

Wautrecht J C, Vandenbossche J L, Englert M

出版信息

Eur Heart J. 1984 May;5(5):404-11. doi: 10.1093/oxfordjournals.eurheartj.a061674.

Abstract

In order to assess the value of pulsed Doppler echocardiography in detection of valvular regurgitation, 63 patients were evaluated for aortic and/or mitral regurgitation using pulsed Doppler echocardiography and selective cineangiography. The Doppler study was considered as positive when a turbulent flow was detected below the aortic valve for aortic insufficiency and behind the mitral valve for mitral insufficiency on a graphic display (time interval histogram) when technically adequate and/or on an audiosignal. These results were compared with standard angiographic evaluation of the regurgitation: pulsed Doppler echocardiography had 94% sensitivity and the specificity rate was very high (87.5%) even for mild regurgitation. Thus, Doppler technique is highly specific and sensitive in detection of aortic and mitral regurgitation when both audiosignal and time-interval histogram are simultaneously performed.

摘要

为了评估脉冲多普勒超声心动图在检测瓣膜反流中的价值,对63例患者进行了评估,使用脉冲多普勒超声心动图和选择性心血管造影术检测主动脉瓣和/或二尖瓣反流。当在图形显示(时间间隔直方图)上,在技术条件充分时,于主动脉瓣下方检测到湍流信号提示主动脉瓣关闭不全,于二尖瓣后方检测到湍流信号提示二尖瓣关闭不全,和/或检测到音频信号时,多普勒检查结果被视为阳性。将这些结果与反流的标准血管造影评估结果进行比较:脉冲多普勒超声心动图的敏感性为94%,即使对于轻度反流,特异性率也非常高(87.5%)。因此,当同时进行音频信号和时间间隔直方图分析时,多普勒技术在检测主动脉瓣和二尖瓣反流方面具有高度的特异性和敏感性。