Gibney R T, Wilson R S, Pontoppidan H
Chest. 1982 Dec;82(6):692-5. doi: 10.1378/chest.82.6.692.
Using on-line computer analysis, the work of breathing (WOB) was measured in ten normal subjects breathing through four continuous positive airway pressure (CPAP) circuits. Two systems (Emerson ventilator) and MGH elastic loaded reservoir bag (ELRB) were high continuous flow systems and two (Puritan-Bennett MA-2 and Bourns Bear Ventilators) were demand valve circuits. Mean values for the WOB were similar with either of the two continuous high flow systems (Emerson 0.41 +/- .21 kg . m . L-1; ERLB 0.49 +/- .29 kg . m . min-1) and significantly lower than with the two demand valve systems (MA-2 0.89 +/- .40 kg . m . min-1; Bourns .71 +/- .30 kg . m . min-1).
通过在线计算机分析,对10名正常受试者通过四个持续气道正压通气(CPAP)回路呼吸时的呼吸功(WOB)进行了测量。两个系统(爱默生呼吸机)和麻省总医院弹性负载储气囊(ELRB)是高持续流量系统,另外两个(伟康MA - 2和博恩斯熊牌呼吸机)是按需阀回路。两种持续高流量系统中的任何一种的WOB平均值相似(爱默生0.41±0.21千克·米·升⁻¹;ELRB 0.49±0.29千克·米·分钟⁻¹),且显著低于两种按需阀系统(MA - 2 0.89±0.40千克·米·分钟⁻¹;博恩斯0.71±0.30千克·米·分钟⁻¹)。