Abstract

Transcutaneous pCO2 electrode response time was optimized by use of a new electrode filling solution composed of NaCl/NaHCO3 electrolyte buffer (100 and 20 mmol/L, respectively) in an equivolume mixture of glycerol and water. The 95% response time to a step change in pCO2 was 49.9 +/- 2.8 s (mean +/- SD) when there was no spacer between the membrane and glass of the electrode. Use of this filling solution during monitoring of severely ill premature infants with cardiopulmonary disease allowed identification of large, unpredictable transient changes in transcutaneous pCO2, and therefore presumably in arterial pCO2, that occurred during capillary blood gas sampling. The changes, which occurred in 19 of 20 samplings, ranged from -1.06 + 2.53 kPa (-8 to +19 Torr). The maximum relative change observed was +29%. These results indicate that the standard protocol for capillary blood collection induces significant transient fluctuations in blood gas tensions. We believe these fluctuations decrease the reliability of capillary pCO2 values for use in clinical management in patient populations similar to ours.

摘要

通过使用一种新的电极填充溶液优化经皮二氧化碳电极的响应时间,该填充溶液由氯化钠/碳酸氢钠电解质缓冲液(分别为100和20 mmol/L)在甘油和水的等体积混合物中组成。当电极的膜与玻璃之间没有间隔物时,对二氧化碳分压阶跃变化的95%响应时间为49.9±2.8秒(平均值±标准差)。在监测患有心肺疾病的重症早产儿期间使用这种填充溶液,能够识别出在毛细血管血气采样期间经皮二氧化碳分压出现的大的、不可预测的瞬时变化,因此推测动脉二氧化碳分压也会出现这种变化。这些变化在20次采样中的19次出现,范围为-1.06 + 2.53 kPa(-8至+19 Torr)。观察到的最大相对变化为+29%。这些结果表明,标准的毛细血管采血方案会在血气张力中引起显著的瞬时波动。我们认为,这些波动降低了在与我们类似的患者群体中用于临床管理的毛细血管二氧化碳分压值的可靠性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验