Longhurst D M, Gooch W M, Castillo R A
Transfusion. 1983 Mar-Apr;23(2):170-2. doi: 10.1046/j.1537-2995.1983.23283172861.x.
Use of a pediatric stainless steel microaggregate screen filter to filter whole blood, gravity-sedimented packed cells, and centrifuged packed cells (cPC) from 20 healthy adults resulted in excessive levels of hemolysis (to 1477.9 mg/dl plasma hemoglobin with 5-day-old cPC) and unacceptable frequency of occlusion (10 to 40 percent).