Suppr超能文献

镓熔点标准:其在临床实验室温度测量中的应用与评估。

The gallium melting-point standard: its application and evaluation for temperature measurements in the clinical laboratory.

作者信息

Bowers G N, Inman S R

出版信息

Clin Chem. 1977;23(4):733-7.

PMID:844169
Abstract

We are impressed with the ease and certainty of calibration electronic thermometers with thermistor probes to +/- 0.01 degree C at the gallium melting point, 29.771(4) degrees C. The IFCC reference method for measuring aspartate aminotransferase activity in serum was run at the reaction temperature of 29.771(4) degrees C. By constantly referencing to gallium as an integral part of the assay procedure, we determined the absolute reaction temperature to IPTS-68 (International Practical Temperature Scale of 1968) to +/- 0.02 degrees C. This unique temperature calibration standard near the center of the range of temperatures commonly used in the clinical laboratory is a valuable addition and can be expected to improve the accuracy of measurements, especially in clinical enzymology.

摘要

我们对采用热敏电阻探头的电子温度计在镓熔点(29.771(4)摄氏度)下轻松且准确地校准至±0.01摄氏度印象深刻。国际临床化学和检验医学联合会(IFCC)测定血清中天冬氨酸转氨酶活性的参考方法是在29.771(4)摄氏度的反应温度下进行的。通过在测定过程中始终将镓作为一个组成部分进行参照,我们将绝对反应温度确定为相对于IPTS - 68(1968年国际实用温标)±0.02摄氏度。这个接近临床实验室常用温度范围中心的独特温度校准标准是一个有价值的补充,有望提高测量的准确性,尤其是在临床酶学方面。