Mayer G S, Shoup R E
J Chromatogr. 1983 Jan 21;255:533-44. doi: 10.1016/s0021-9673(01)88308-4.
To enhance the selectivity of liquid chromatographic-electrochemical assays for biogenic amines and metabolites in brain tissue, multiple electrode transducers were investigated. Two configurations of dual working electrodes were examined: parallel-adjacent and series arrangements. Using the raw detector currents with each configuration, peak-height ratios from simultaneously generated chromatograms were calculated to assess the selectivity of the instrument for direct injections of brain tissue supernatant. Ratios were consistent with injections of standards. Nearly coeluting peaks such as norepinephrine and 3-methoxy-4-hydroxyphenylglycol were resolved by using dual detector electrodes in series; only the catecholamines were detected at the downstream electrode owing to their electrochemical reversibility. The scheme was applicable to the assay of norepinephrine, 3,4-dihydroxyphenylacetic acid, dopamine, homovanillic acid, serotonin, 5-hydroxytryptophan and 5-hydroxyindole-3-acetic acid in brain tissue in less than 15 min.
为提高液相色谱 - 电化学分析法对脑组织中生物胺及其代谢物的选择性,对多电极换能器进行了研究。考察了两种双工作电极配置:平行相邻和串联排列。使用每种配置下的原始检测器电流,计算同时生成的色谱图的峰高比,以评估该仪器对直接注射脑组织上清液的选择性。这些比值与标准品注射结果一致。通过串联使用双检测电极,可分辨出几乎同时洗脱的峰,如去甲肾上腺素和3 - 甲氧基 - 4 - 羟基苯乙二醇;由于儿茶酚胺具有电化学可逆性,仅在下游电极检测到它们。该方法适用于在不到15分钟内测定脑组织中的去甲肾上腺素、3,4 - 二羟基苯乙酸、多巴胺、高香草酸、血清素、5 - 羟色氨酸和5 - 羟基吲哚 - 3 - 乙酸。