Gordon M A, Duncan R A, Kingsley L C
J Clin Microbiol. 1981 Feb;13(2):283-5. doi: 10.1128/jcm.13.2.283-285.1981.
A soluble-antigen fluorescent-antibody technique for serodiagnosis of toxoplasmosis, employing a special fluorometer (FIAX; International Diagnostic Technology, Santa Clara, Calif.), was evaluated in comparison with a standard indirect fluorescent-antibody method which uses a fluorescence microscope. Titers obtained by the new method on 324 serum samples correlated very well with those of the conventional procedure. There was good intralaboratory and interlaboratory agreement in runs on selected samples. The FIAX method is simple, rapid, and reliable and has the advantage of objectivity.
一种用于弓形虫病血清诊断的可溶性抗原荧光抗体技术,采用一种特殊的荧光计(FIAX;国际诊断技术公司,加利福尼亚州圣克拉拉),与使用荧光显微镜的标准间接荧光抗体方法相比进行了评估。通过新方法对324份血清样本测得的滴度与传统方法测得的滴度相关性非常好。在对选定样本的检测中,实验室内部和实验室之间的结果一致性良好。FIAX方法简单、快速且可靠,具有客观性的优点。