Oliver D G, Sanders A H, Hogg R D, Hellman J W
J Immunol Methods. 1981;42(2):195-201. doi: 10.1016/0022-1759(81)90149-6.
Temperature studies of microtitration plates demonstrate that the use of a common bacteriology incubator for heating the plates can cause a phase log of over 30 min for the fluid in the wells to reach 37 degrees C from ambient temperature, and that a temperature gradient of as much as 1.6 degree C can exist between the peripheral and center wells. This gradient is a cause of the "rim" of edge effect noted in enzyme immunoassay using microtitration plates. The problem is corrected by the use of a specially designed forced air microtitration plate incubator.
微量滴定板的温度研究表明,使用普通细菌培养箱加热滴定板会导致孔内液体从室温达到37摄氏度的阶段滞后超过30分钟,并且外周孔和中心孔之间可能存在高达1.6摄氏度的温度梯度。这种梯度是使用微量滴定板的酶免疫测定中出现边缘效应“边缘”的原因。使用专门设计的强制空气微量滴定板培养箱可纠正该问题。