Relative efficacy of an incubator and an open warmer in producing thermoneutrality for the small premature infant.

作者信息

LeBlanc M H

出版信息

Pediatrics. 1982 Apr;69(4):439-45.

PMID:7070890
Abstract

To determine which warming system more closely approximates a neutral thermal environment, the oxygen consumption of 16 premature babies, weighing less than 1,500 gm each, was measured in a convectively heated incubator and under an open radiant warmer. Both systems were controlled to maintain a skin temperature of 36 C. The oxygen consumption of the infants was significantly higher under the radiant warmer. This result is in agreement with published data for this group of premature infants.

摘要

为了确定哪种保暖系统更接近中性热环境,对16名体重均不足1500克的早产婴儿在对流加热培养箱和开放式辐射保暖器下的耗氧量进行了测量。两种系统均被控制以维持皮肤温度为36摄氏度。婴儿在辐射保暖器下的耗氧量显著更高。这一结果与针对这组早产婴儿的已发表数据一致。