Decontamination of laboratory microbiological waste by steam sterilization.

作者信息

Rutala W A, Stiegel M M, Sarubbi F A

出版信息

Appl Environ Microbiol. 1982 Jun;43(6):1311-6. doi: 10.1128/aem.43.6.1311-1316.1982.

Abstract

A steam sterilizer (autoclave) was tested to determine the operating parameters that affected sterilization of microbiological waste. Tests involved standardized loads (5, 10 ad 15 lb [ca. 2.27, 4.54, and 6.80 kg, respectively]) contaminated petri plates in autoclave bags placed in polypropylene or stainless steel containers. Thermal and biological data were obtained by using a digital potentiometer and a biological indicator containing spores of Bacillus stearothermophilus, respectively. The transfer of heat was more efficient when smaller loads of microbiological waste were tested and stainless steel rather than polypropylene containers were used. A single bag with the sides rolled down to expose the top layer of petri plates allowed heat to pass better than did a single bag with the top constricted by a twist-tie. The presence of water in the autoclave bag did not significantly improve heat-up time in stainless steel or polypropylene containers. The results of biological tests substantiated the temperature data. When 10 or 15 lb of microbiological waste was exposed to various test conditions, the only condition that ensured the destruction of B. stearothermophilus involved the use of a stainless steel container (with or without water) for 90 min. Autoclaving for 45 min resulted in the destruction of bacteria included in 10 lb (136 +/- 3 plates) or 15 lb (205 +/- 6 plates) of microbiological waste when stainless steel containers with or without water or polypropylene containers with water used, whereas 60 min was required to kill all bacteria if polypropylene containers without water were used.

摘要

对一台蒸汽灭菌器(高压灭菌锅)进行了测试,以确定影响微生物废物灭菌效果的运行参数。测试涉及标准负载(分别为5磅、10磅和15磅[约2.27千克、4.54千克和6.80千克]),将受污染的培养皿置于高压灭菌袋中,再把袋子放入聚丙烯或不锈钢容器内。分别使用数字电位计和含有嗜热脂肪芽孢杆菌孢子的生物指示剂获取热学和生物学数据。当测试较小负载的微生物废物且使用不锈钢而非聚丙烯容器时,热传递效率更高。一个将袋边向下卷以露出培养皿顶层的袋子比一个用束带收紧顶部的袋子传热效果更好。高压灭菌袋中有水并不会显著缩短不锈钢或聚丙烯容器的升温时间。生物学测试结果证实了温度数据。当10磅或15磅微生物废物暴露于各种测试条件下时,确保嗜热脂肪芽孢杆菌被灭活的唯一条件是使用不锈钢容器(有水或无水)90分钟。当使用有水或无水的不锈钢容器或有水的聚丙烯容器时,高压灭菌45分钟可使10磅(136±3个培养皿)或15磅(205±6个培养皿)微生物废物中的细菌被灭活;而如果使用无水的聚丙烯容器,则需要60分钟才能杀死所有细菌。

相似文献

2
Decontaminating infectious laboratory waste by autoclaving.通过高压灭菌对传染性实验室废物进行消毒。
Appl Environ Microbiol. 1982 Sep;44(3):690-4. doi: 10.1128/aem.44.3.690-694.1982.
3
Effects of steam sterilization on the contents of sharps containers.蒸汽灭菌对锐器盒内物品的影响。
Am J Infect Control. 1993 Feb;21(1):28-33. doi: 10.1016/0196-6553(93)90204-h.
10
Does incineration turn infectious waste aseptic?
J Hosp Infect. 2005 Aug;60(4):304-6. doi: 10.1016/j.jhin.2005.01.016.

本文引用的文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验