Weiss G H, Talbert A J, Brooks R A
Phys Med Biol. 1982 Sep;27(9):1151-62. doi: 10.1088/0031-9155/27/9/005.
An artefact that can occur in CT scans with too few views is the appearance of streaks at some distance from the sharp edge that causes them. We investigate, for parallel beam scanners, the effectiveness of the method of phantom views, as well as other methods of angular interpolation, in diminishing the view-limited streak artefact. The phantom view method consists of inserting data points obtained by interpolation in angle, along with the real data points. We show by a theoretical analysis, as well as by using simulated scans, that the phantom view method does tend to diminish the distant streaks more than other methods tested. Furthermore, we show that there is a point of diminishing returns in increasing the number of phantom views.
在视图过少的CT扫描中可能出现的一种伪影是,在距产生伪影的尖锐边缘一定距离处出现条纹。对于平行束扫描仪,我们研究了虚拟视图方法以及其他角度插值方法在减少视图受限条纹伪影方面的有效性。虚拟视图方法包括插入通过角度插值获得的数据点以及真实数据点。我们通过理论分析以及使用模拟扫描表明,虚拟视图方法确实比其他测试方法更倾向于减少远处的条纹。此外,我们表明增加虚拟视图的数量存在收益递减点。