Class II (laminar flow) biological safety cabinet.

作者信息

Newsom S W

出版信息

J Clin Pathol. 1979 May;32(5):505-13. doi: 10.1136/jcp.32.5.505.

Abstract

A microbiological survey of the effectiveness of class II (laminar flow) 'safety' cabinets found in the UK in the last six years is recorded. Only two of the nine units tested approached the containment of aerosols achieved by a good class I (exhaust protective) cabinet. The others were potentially hazardous if used with pathogenic material. The National Sanitation Foundation and the British Standards Institute have now laid down adequate specifications based on biological containment, and hopefully the conforming cabinets will be much better; even so, the purchase of a cabinet must be undertaken with care, and the cabinet requires frequent monitoring during use and after servicing.

摘要

记录了对过去六年在英国发现的二级(层流)“安全”柜有效性的微生物学调查。测试的九个装置中只有两个达到了一级(排气防护)优质柜所实现的气溶胶遏制效果。其他的如果与致病物质一起使用可能具有危险性。国家卫生基金会和英国标准协会现已根据生物遏制制定了适当的规范,希望符合标准的柜子会好得多;即便如此,购买柜子时必须谨慎,并且柜子在使用期间和维修后需要经常监测。