Thévenot D R, Sternberg R, Coulet P
Diabetes Care. 1982 May-Jun;5(3):203-6. doi: 10.2337/diacare.5.3.203.
A very sensitive glucose electrode was developed using a glucose-oxidase membrane and an anodically polarized platinum disk. Calibration curves were linear over 4.5 concentration decades. It was adapted for human blood samples. The lifetime of glucose-oxidase collagen membranes was greater than 3 yr at 4 degrees C and was approximately 6 mo at 20-30 degrees C.
利用葡萄糖氧化酶膜和阳极极化铂盘研制出了一种非常灵敏的葡萄糖电极。校准曲线在4.5个浓度十进位数范围内呈线性。它适用于人体血液样本。葡萄糖氧化酶胶原膜在4℃下的寿命大于3年,在20 - 30℃下约为6个月。