Case R V, Mezei L M
Clin Chem. 1978 Dec;24(12):2145-50.
An enzymatic, radiochemical assay procedure for measuring serum gentamicin by use of gentamicin 3-acetyl-transferase and [acetyl-14C]acetyl-Coenzyme A is described and evaluated. The enzyme stoichiometrically and quantitatively transfers a radioactive label to the analyte during a 10-min incubation at 37 degrees C. The labeled gentamicin is then absorbed onto phosphocellulose paper discs, which are washed to remove unreacted [acetyl-14C]-acetyl-Coenzyme A and counted in a liquid scintillation counter for 1 min each. The assay detects gentamicin in concentrations as low as 0.2 mg/liter and gives a linear response to concentrations as high as 20 mg/liter. Sisomicin, a structural analog of gentamicin, is measured by the procedure, and tobramycin and netilmicin are slightly reactive. No other interferents were found among other aminoglycosides, other antibiotics, or substances endogenous to serum. Results by the new method are compared to those by radioimmunoassay and a microbiological method.
描述并评估了一种使用庆大霉素3 - 乙酰转移酶和[乙酰 - 14C]乙酰辅酶A来测定血清中庆大霉素的酶促放射化学分析方法。在37℃孵育10分钟期间,该酶以化学计量的方式将放射性标记定量转移至分析物上。然后将标记的庆大霉素吸附到磷酸纤维素纸圆盘上,冲洗这些圆盘以去除未反应的[乙酰 - 14C] - 乙酰辅酶A,并在液体闪烁计数器中各计数1分钟。该分析方法可检测低至0.2毫克/升浓度的庆大霉素,并且对高达20毫克/升的浓度呈线性响应。该方法可测定庆大霉素的结构类似物西索米星,而妥布霉素和奈替米星有轻微反应。在其他氨基糖苷类、其他抗生素或血清内源性物质中未发现其他干扰物。将新方法的结果与放射免疫分析方法和微生物学方法的结果进行了比较。