Keller F, Feldmann K, Metz F J
J Clin Chem Clin Biochem. 1981 Mar;19(3):169-71.
The estimation of serum albumin obtained with SMAR 12/60 by the bromcresol green method are in good agreement with values determined by electrophoresis only in the normal range. Interestingly marked discrepancies occur in the clinically relevant cases of hypalbuminosis. The shortening of the reaction time between bromcresol green and serum results in a better agreement between the values obtained with SMAR 12/60 and the values determined by electrophoresis over the whole range.
使用SMAR 12/60通过溴甲酚绿法测定的血清白蛋白值仅在正常范围内与通过电泳测定的值高度一致。有趣的是,在临床上相关的低白蛋白血症病例中出现了明显差异。缩短溴甲酚绿与血清之间的反应时间,使得在整个范围内使用SMAR 12/60获得的值与通过电泳测定的值之间的一致性更好。