小儿外科手术期间的经皮氧分压监测

Transcutaneous PO2 monitoring during pediatric surgery.

作者信息

Venus B, Patel K C, Pratap K S, Konchigeri H, Vidyasagar D

出版信息

Crit Care Med. 1981 Oct;9(10):714-6. doi: 10.1097/00003246-198110000-00009.

Abstract

The value of continuous transcutaneous oxygen (PtcO2) monitoring in the presence of halothane and nitrous oxide (N2O) was studied in 10 infants. All infants were continuously monitored during the study with a Clark-type skin electrode with Mylar membrane. The polarization voltage of the membrane was changed to 600 mV to make it insensitive to N2O and halothane. The accuracy of PtcO2 electrode was compared with simultaneously measured PaO2 at different intervals. Correlation of PaO2 with PtcO2 was sought during hypoxic-normoxic state (PaO2 between 27--92 torr) and hyperoxic state (PaO2 between 105--439 torr). During hypoxic-normoxic state, the correlation between PaO2 and PtcO2 values was 0.94. During the hyperoxic state, poor correlation existed between PaO2 and PtcO2 (r = 0.51). Although PtcO2 did not correlate with PaO2 during hyperoxia, it consistently overestimated PaO2 and thereby, provided a predictive ability by over-diagnosing hyperoxia. When combined with continuous monitoring of inspired oxygen tension to maintain normoxia, continuous monitoring of PtcO2 will reduce the frequency of PaO2 analysis and improve patient care during anesthesia.

摘要

在10名婴儿中研究了在氟烷和氧化亚氮(N2O)存在的情况下持续经皮氧分压(PtcO2)监测的价值。在研究期间,所有婴儿均使用带有聚酯薄膜的Clark型皮肤电极进行持续监测。将薄膜的极化电压改为600 mV,使其对N2O和氟烷不敏感。在不同时间间隔将PtcO2电极的准确性与同时测量的动脉血氧分压(PaO2)进行比较。在低氧-正常氧状态(PaO2在27-92托之间)和高氧状态(PaO2在105-439托之间)下寻找PaO2与PtcO2之间的相关性。在低氧-正常氧状态下,PaO2与PtcO2值之间的相关性为0.94。在高氧状态下,PaO2与PtcO2之间存在较差的相关性(r = 0.51)。尽管在高氧期间PtcO2与PaO2不相关,但它始终高估PaO2,从而通过过度诊断高氧提供了预测能力。当与持续监测吸入氧分压以维持正常氧合相结合时,持续监测PtcO2将减少PaO分析的频率并改善麻醉期间的患者护理。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验