Xu C X, Tang X J, Niu Z, Li Z M
Int J Artif Organs. 1981 Jul;4(4):200-4.
A new method for the preparation of the cross-linked agarose beads entrapped activated charcoal (CAAC) is reported. Since the agarose-encapsulated adsorbents reported elsewhere cannot stand high temperature for sterilization, the CAAC has the advantage of thermal stability to withstand autoclave at 121 degrees C. for 1/2 hour without breaking up or melting. A further advantage of CAAC is that the adsorbent has a much better consistency with good mechanical strength and elasticity, so that it can be formed into beads of a diameter less than 1 mm. This will not only give a better adsorption capacity than larger beads, but can also assure a better blood flow than soft beads which usually interfere in hemoperfusion due to compacting and sludging. Preliminary investigations indicate that the CAAC is relatively hemocompatible.
报道了一种制备交联琼脂糖珠包埋活性炭(CAAC)的新方法。由于其他地方报道的琼脂糖包封吸附剂不能承受高温灭菌,CAAC具有热稳定性优势,能在121摄氏度下经受高压灭菌半小时而不破碎或熔化。CAAC的另一个优点是吸附剂具有更好的一致性,具备良好的机械强度和弹性,因此可以制成直径小于1毫米的珠子。这不仅比大珠子具有更好的吸附能力,而且比通常因压实和淤塞而干扰血液灌流的软珠子能确保更好的血流。初步研究表明,CAAC具有相对较好的血液相容性。