Heinemann G, Günner H
Blut. 1980 Apr;40(4):275-83. doi: 10.1007/BF01080187.
Platelet counts in whole blood were performed by means of a new particle analyzer with a hydrodynamic focused aperture (Ultra-Flo 100) and its values were compared with the results obtained by an optical counting chamber. The comparison yielded good agreement of the values within the evaluated range of 3-963 G/l thrombocytes. Especially remarkable is the good correlation of the markedly thrombocytopenic values. Equally reliable is the counting of thrombocytes in blood with markedly elevated erythrocytes. The simple procedure of sample preparing combined with the stand-by technic of the analyzer provides advantageous conditions for using it likewisely in routine and emergency laboratories.
采用带有流体动力聚焦孔的新型颗粒分析仪(Ultra-Flo 100)对全血中的血小板进行计数,并将其结果与光学计数室获得的结果进行比较。在3 - 963 G/l血小板的评估范围内,比较结果显示数值具有良好的一致性。尤其值得注意的是,在血小板明显减少的数值方面具有良好的相关性。对于红细胞明显升高的血液中的血小板计数同样可靠。样品制备的简单程序与分析仪的备用技术相结合,为在常规和急诊实验室中同样使用该分析仪提供了有利条件。