Catheter tip pH electrodes for continuous intravascular recording.

作者信息

Cobbe S M, Poole-Wilson P A

出版信息

J Med Eng Technol. 1980 May;4(3):122-4. doi: 10.3109/03091908009161104.

Abstract

A small, flexible catheter-tip pH electrode has been developed for continuous measurement of intravascular pH in man and animals. The electrode utilises a pH sensitive polymer membrane rather than pH glass. The electrode passes through the lumen of a conventional cardiac catheter, by which it may be introduced to any desired site in the circulation. The electrode has no cross-sensitivity to other ions, to osmolality or oxygen tension. It is insensitive to flow and pressure. The response time is less than 300 ms, and the mean drift in vivo is 0.01 pH units/h. Continuous recordings of intravascular pH in animals have been made for up to 7h. The electrodes can be sterilised, and have been used in man.

摘要

已开发出一种小型、灵活的导管尖端pH电极,用于连续测量人和动物的血管内pH值。该电极使用对pH敏感的聚合物膜而非pH玻璃。该电极可穿过传统心导管的内腔,借此可将其引入循环系统中的任何所需部位。该电极对其他离子、渗透压或氧分压无交叉敏感性。它对流量和压力不敏感。响应时间小于300毫秒,体内平均漂移为0.01pH单位/小时。已对动物进行了长达7小时的血管内pH值连续记录。这些电极可进行消毒,并已在人体中使用。