Continuous 24 hour ambulatory monitoring of intragastric pH in man.

作者信息

Kapur B K, Howlett P J, Kenyon N G, Lunt M J, Mills J G, Smallwood R H, Wilson A J, Bardhan K D

出版信息

Clin Phys Physiol Meas. 1987 May;8(2):123-32. doi: 10.1088/0143-0815/8/2/003.

Abstract

A system has been developed which permits continuous 24 hour ambulatory recording of intragastric pH under near-physiological conditions. The system utilises a Cecar combination pH electrode connected through a pre-amplifier to an Oxford Medical Systems Medilog 4-24 cassette recorder, and the pH recorded continuously on to tape. The data is replayed at high speed, digitised and then analysed. Sources of error included system drift, system noise and the effect of changes in electrode environment due to saliva, food, loss of fluid contact, temperature variations and electrode time constant. These were found to be small. In contrast, the major changes in pH detected reflect real changes at the electrode tip. These changes can be recorded by the system. Developments in electronics and improvements in both the size and quality of pH electrodes, over the past few decades, has enabled intragastric pH recordings to be made with greater ease. However, a large number of investigators still assess gastric acidity by infrequent, intermittent sampling of gastric secretion through a naso-gastric tube. This new ambulatory system is a significant improvement and its versatility allows studies in acid secretion physiology, pharmacology and pathophysiology.