Goldstein S R, Peterson J I, Fitzgerald R V
J Biomech Eng. 1980 May;102(2):141-6. doi: 10.1115/1.3138210.
A flexible 0.4-mm-dia pH probe potentially suitable for physiological use has been developed. It is based on the concept of utilizing two single plastic fiber optic strands to illuminate and remotely sense the color change of a dye indicator contained within an acutely implanted sealed cellulosic hollow fiber permeable to hydrogen ions. A supporting electronic module provides tungsten filament illumination, light sensing with a photodiode/operational amplifier, analog and digital circuitry to provide appropriate signal averaging and processing, and a mechanical assembly to enable the optical density measurements to be made both at 560 nm and, for normalization purposes, in the red. Over the physiological pH range from 7.0 to 7.4, the fiber optic probe agrees with a standard glass pH electrode to within 0.01 pH units in buffer solutions, to within 0.017 pH units in heparinized dog blood in vitro, and it has performed successfully while implanted in the jugular vein of a sheep.
一种直径为0.4毫米的柔性pH探头已被开发出来,它可能适用于生理学用途。它基于这样一个概念,即利用两根单根塑料光纤束来照亮并远程感知急性植入的、对氢离子可渗透的密封纤维素中空纤维内所含染料指示剂的颜色变化。一个支持电子模块提供钨丝照明、用光电二极管/运算放大器进行光感测、用于提供适当信号平均和处理的模拟和数字电路,以及一个机械组件,以便能够在560纳米处进行光密度测量,并且为了归一化目的,还能在红色波段进行测量。在7.0至7.4的生理pH范围内,该光纤探头在缓冲溶液中与标准玻璃pH电极的误差在0.01 pH单位以内,在体外肝素化犬血中误差在0.017 pH单位以内,并且在植入绵羊颈静脉时已成功运行。