Determination of functional residual capacity with 133-xenon radiospirometry. Comparison with body plethysmography and helium spirometry. Effect of body position.

作者信息

Kauppinen-Walin K, Sovijärvi A R, Muittari A, Uusitalo A

出版信息

Scand J Clin Lab Invest. 1980 Jun;40(4):347-54. doi: 10.3109/00365518009092654.

Abstract

This study was undertaken to estimate the accuracy of 133-xenon radiospirometry for determination of FRC in healthy subjects. Forty healthy volunteers, both smokers and non-smokers, were examined. The FRC of each subject was concurrently determined with radiospirometric, He-dilution in closed circuit, and body plethysmographic methods. The radiospirometric and He-dilution measurements were done in supine and in sitting positions, the body plethysmography on sitting subjects, only. The mean FRC measured by radiospirometry (FRCRS) was 0.72 1 larger than that measured by helium spirometry (FRCHe) in sitting position (P < 0.01). In supine position the FRCRS was 0.65 1 larger than the FRCHe (P < 0.01). The body plethysmography gave FRC (TGV) 0.35 1 larger than the FRCHe sitting (P < 0.01). The FRCHe and the FRCRS in the sitting position were 0.48 and 0.55 1 larger than in the supine position (P < 0.01), respectively. Trapped air correlated significantly (P < 0.01) with the difference FRCRS-FRCHe, when sitting. The results indicate that the FRC determined radiospirometrically is significantly larger than the FRC determined with He-spirometry. The difference is systematic, suggesting that it is caused by 133-xenon dissolved in blood and accumulated in tissues of the thoracic cage and by dissimilar representation of trapped air in FRCRS and FRCHe. With eventual correction of the systematic error, the FRC obtained as a by-product of radiospirometry may be used, e.g. for clinical purposes.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验