Lacombe F, Cazaux N, Briais A, Labroille G, Puntous M, Reiffers J, Belloc F, Boisseau M R, Bernard P
Laboratoire d'Hématologie, Hôpital Haut-Lévèque, CHU Bordeaux, France.
Am J Clin Pathol. 1995 Nov;104(5):495-502. doi: 10.1093/ajcp/104.5.495.
To evaluate the leukocyte differential flags of the Cobas Argos 5 Diff., the authors performed a comparative study between their current analyzer, the Technicon H2, and the manual leukocyte differential. Samples (n = 1,600) were collected from the Blood Disease Department of their hospital and were tested on both Cobas Argos 5 Diff. (ABX/Roche Hematology Division, Montpellier, France) and Technicon H2. Abnormalities of the manual leukocyte differential (immature granulocytes, blast cells, atypical lymphocytes, hyperbasophil cells, erythroblasts, and hairy cells) were found in 597 samples. The authors determined the best cut-off of the quantitative flags--atypical lymphocytes (ALYs) and large immature cells (LICs)--using the likelihood ratio method, and the capability of the 5 Diff. qualitative flags to determine abnormal subpopulations by the predictive value of a positive result. The presence of particular combinations of flags was associated with band cells and blast cells of acute lymphoblastic leukemia (ALL) or acute myeloblastic leukemia (AML). The sensitivity, specificity, and efficiency of the two analyzers were 92.1, 53.9, and 68.2, respectively, for the Argos 5 Diff., and 93.5, 36.7 and 57.9, respectively, for Technicon H2. In conclusion, the Cobas Argos 5 Diff. is well adapted for detecting leukocyte abnormalities.
为评估Cobas Argos 5 Diff.的白细胞分类标记,作者在他们现有的分析仪Technicon H2和手工白细胞分类之间进行了一项比较研究。从他们医院的血液科收集了样本(n = 1600),并在Cobas Argos 5 Diff.(ABX/罗氏血液学部,法国蒙彼利埃)和Technicon H2上进行检测。在597个样本中发现了手工白细胞分类的异常(未成熟粒细胞、原始细胞、非典型淋巴细胞、嗜碱性粒细胞增多、有核红细胞和毛细胞)。作者使用似然比方法确定了定量标记——非典型淋巴细胞(ALYs)和大未成熟细胞(LICs)的最佳临界值,以及5 Diff.定性标记通过阳性结果预测值确定异常亚群的能力。特定标记组合的存在与急性淋巴细胞白血病(ALL)或急性髓细胞白血病(AML)的杆状核细胞和原始细胞有关。对于Argos 5 Diff.,两种分析仪的敏感性、特异性和效率分别为92.1、53.9和68.2,对于Technicon H2分别为93.5、36.7和57.9。总之,Cobas Argos 5 Diff.非常适合检测白细胞异常。