成人医院中BACTEC 9240血培养系统与BacT/Alert血培养系统的比较

Comparison of BACTEC 9240 and BacT/Alert blood culture systems in an adult hospital.

作者信息

Smith J A, Bryce E A, Ngui-Yen J H, Roberts F J

机构信息

Department of Pathology, Vancouver Hospital and Health Sciences Centre, British Columbia, Canada.

出版信息

J Clin Microbiol. 1995 Jul;33(7):1905-8. doi: 10.1128/jcm.33.7.1905-1908.1995.

Abstract

In a study comparing the BACTEC 9240 (Becton Dickinson, Canada Inc., Mississauga, Ontario, Canada) and the BacT/Alert (Organon Teknika, Scarborough, Ontario, Canada) blood culture systems, 6,437 sets of four bottles of 10,412 cultures submitted were deemed acceptable for evaluation. There were 586 clinically significant isolates from 544 positive blood cultures. Of the 544 positive blood cultures, 329 were positive in both systems, 157 were positive in the BACTEC 9240 system alone, and 58 were positive in the BacT/Alert system alone (P < 0.001). These cultures represented 421 bacteremic episodes, of which 290 were detected by both systems, 84 were detected by the BACTEC 9240 system alone, and 47 were detected by the BacT/Alert system alone (P < 0.001). The difference between the two systems in terms of microbial recovery was attributable largely to differences in the ability to grow staphylococci. Of 82 monomicrobial isolates of Staphylococcus aureus, 46 were isolated by both systems, 30 were isolated by the BACTEC 9240 system alone, and 6 were isolated by the BacT/Alert system alone (P < 0.0001). A similar pattern was noted in the case of coagulase-negative staphylococci, of which 48 were isolated in both systems, 46 were isolated in the BACTEC 9240 system alone, and 5 were isolated in the BacT/Alert system alone (P < 0.0001). The false-positive rate in the BACTEC 9240 system was 0.5%, and that in the BacT/Alert system was 1.2%. Both systems provided satisfactory service mechanically, and in terms of computer software and costs, both systems were similar. The differences in the recovery of organisms, notably staphylococci, may be due in part to the fact that in the present study the BacT/Alert medium did not contain antimicrobial agent-removing devices, whereas the BACTEC 9240 medium did. Until this difference is resolved, we consider the BACTEC 9240 system to more adequately meet our hospital's requirements.

摘要

在一项比较BACTEC 9240(加拿大安大略省密西沙加市百特迪金森公司)和BacT/Alert(加拿大安大略省士嘉堡市奥加农泰克尼卡公司)血培养系统的研究中,提交的6437套共10412瓶培养物被认为可用于评估。544份阳性血培养中有586株具有临床意义的分离株。在这544份阳性血培养中,329份在两个系统中均为阳性,157份仅在BACTEC 9240系统中呈阳性,58份仅在BacT/Alert系统中呈阳性(P<0.001)。这些培养物代表421次菌血症发作,其中290次由两个系统均检测到,84次仅由BACTEC 9240系统检测到,47次仅由BacT/Alert系统检测到(P<0.001)。两个系统在微生物回收率方面的差异很大程度上归因于葡萄球菌生长能力的差异。在82株金黄色葡萄球菌单一微生物分离株中,46株由两个系统均分离到,30株仅由BACTEC 9240系统分离到,6株仅由BacT/Alert系统分离到(P<0.0001)。凝固酶阴性葡萄球菌的情况也有类似模式,其中48株由两个系统均分离到,46株仅由BACTEC 9240系统分离到,5株仅由BacT/Alert系统分离到(P<0.0001)。BACTEC 9240系统的假阳性率为0.5%,BacT/Alert系统的假阳性率为1.2%。两个系统在机械性能、计算机软件和成本方面均提供了令人满意的服务。微生物回收率的差异,尤其是葡萄球菌的差异,可能部分归因于在本研究中BacT/Alert培养基不含抗菌剂去除装置,而BACTEC 9240培养基含有。在这种差异得到解决之前,我们认为BACTEC 9240系统更能充分满足我们医院的需求。

相似文献

引用本文的文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验