Berry S M, Dombrowski M P, Blessed W B, Bichalski J A, Jones T B, Cotton D B
Department of Obstetrics and Gynecology, Wayne State University/Hutzel Hospital, Detroit, Michigan.
Obstet Gynecol. 1993 Mar;81(3):417-20.
To determine whether hemoglobin quantitations using the Hemocue system, a rapid and portable hemoglobin photometer, on fetal blood obtained via funipuncture were accurate compared to the Coulter S-Plus IV. We also examined whether gestational age or extremes in hemoglobin levels significantly affected the accuracy of the Hemocue system.
We performed fetal hemoglobin quantitations using both systems on 58 specimens obtained between 18-38 weeks' gestation. Correlation between values by both systems was determined by linear regression analysis. The effects of gestational age and hemoglobin extremes on the accuracy of the Hemocue system were evaluated by stepwise regression.
The mean Hemocue value was 12.0 +/- 2.4 g/dL (range 3.3-16.4); the mean Coulter value was 11.7 +/- 2.3 g/dL (range 3.6-16.2). The regression equation for Hemocue (y) versus Coulter (x) values was y = 0.72 + 0.97x; r = 0.94 (P < .0001). Neither gestational age nor hemoglobin extremes significantly affected the accuracy of the Hemocue system.
The Hemocue system is rapid and accurate for fetal hemoglobin quantitation between 18-38 weeks' gestation regardless of extremes in gestational age or hemoglobin levels.