Seidel G, Kaps M, Gerriets T, Hutzelmann A
Department of Neurology, Medical University of Lübeck, Germany.
J Neuroimaging. 1995 Apr;5(2):105-8. doi: 10.1111/jon199552105.
Transcranial duplex color-flow sonography provides visualization of intracranial structures, and measures angle-corrected blood flow velocity in the basal cerebral arteries of adults. In 44 patients with central nervous system disease, the oblique diameters of the third and the middle part of the lateral ventricle were measured by transcranial duplex color-flow sonography using a system with a 2.5-MHz transducer, and compared to computed tomography measurements. The correlation coefficients for the third and lateral ventricle measurements were r = 0.83 (p < 0.0001, N = 38) and 0.73 (p < 0.0001, N = 78), respectively. A second investigation was performed by transcranial duplex to evaluate intraobserver and interobserver reproducibility. The correlation coefficient representing interobserver reproducibility in 27 patients was r = 0.87 (p < 0.0001, N = 47) for the lateral ventricle and r = 0.9 (p < 0.0001, N = 49) for the third ventricle. The intraobserver reproducibility correlation coefficient was r = 0.93 (p < 0.0001, N = 22) for the lateral ventricle in 12 patients. In 49 healthy volunteers the oblique diameters of the lateral and third ventricles were age dependent, measuring 16.7 +/- 2.3 mm and 4.8 +/- 1.9 mm, respectively, in those younger than 59 years, compared to 19.0 +/- 2.9 mm and 7.6 +/- 2.1 mm in those 60 years or older. Therefore, transcranial duplex color-flow sonography measures noninvasively the third and the lateral ventricle in adults.
经颅双功彩色血流超声检查可显示颅内结构,并测量成人脑基底动脉的角度校正血流速度。在44例中枢神经系统疾病患者中,使用2.5兆赫换能器系统,通过经颅双功彩色血流超声检查测量侧脑室第三部分和中间部分的斜径,并与计算机断层扫描测量结果进行比较。第三脑室和侧脑室测量的相关系数分别为r = 0.83(p < 0.0001,N = 38)和0.73(p < 0.0001,N = 78)。进行了第二项经颅双功检查以评估观察者内和观察者间的可重复性。27例患者中,代表观察者间侧脑室可重复性的相关系数r = 0.87(p < 0.0001,N = 47),第三脑室为r = 0.9(p < 0.0001,N = 49)。12例患者中,观察者内侧脑室可重复性相关系数r = 0.93(p < 0.0001,N = 22)。在49名健康志愿者中,侧脑室和第三脑室的斜径与年龄有关,59岁以下者分别为16.7±2.3毫米和4.8±1.9毫米,60岁及以上者为19.0±2.9毫米和7.6±2.1毫米。因此,经颅双功彩色血流超声检查可无创测量成人的第三脑室和侧脑室。