Interobserver and interequipment variability of echo-Doppler examination of the portal vein: effect of a cooperative training program.

作者信息

Sabbà C, Merkel C, Zoli M, Ferraioli G, Gaiani S, Sacerdoti D, Bolondi L

机构信息

Istituto di Clinica Medica I, University of Bari, Italy.

出版信息

Hepatology. 1995 Feb;21(2):428-33. doi: 10.1002/hep.1840210225.

Abstract

The aim of this study was to assess the interobserver, interequipment, and time-dependent variabilities of echo-Doppler measurements of portal blood flow velocity (PBV), portal vein diameters (PVDs) and their derived parameters, portal blood flow (PBF), and congestion index (CI) in cirrhotic patients. The influence of a cooperative training program of the operators on the reproductibility of the results was also investigated. The echo-Doppler parameters were independently measured in 15 patients by four skilled operators, using four echo-Doppler machines (Acuson, ATL, Hitachi-Esaote, Toshiba. Eight of the 15 patients were restudied after 15 days by the same operators using only one machine. Significantly different values of PBV, PBF, and CI were obtained. PBV variance was equipment-related (32%) and operator-related in a smaller portion (5%). No systematic effect related to the time of investigation was found. After training to define a precise protocol, new measurements were performed by four operators on 8 different patients. No significant differences were found among the operators for any of the parameters and the 95% confidence limits (CL) and coefficients of variation (CV) of PBV showed a marked decrease (CL from +/- 26.4% to +/- 15.6%). These results indicate that (1) a significant systematic variability exists between Doppler measurements with different equipment; (2) there is no significant time-dependent systematic variability of Doppler measurements; and (3) a cooperative training program reduces the interobserver variability for direct measurements, such as PBV.

摘要

本研究旨在评估肝硬化患者门静脉血流速度(PBV)、门静脉直径(PVD)及其衍生参数门静脉血流量(PBF)和充血指数(CI)的超声多普勒测量结果在观察者间、设备间以及随时间的变异性。同时还研究了操作人员合作培训计划对结果可重复性的影响。由四名熟练操作人员使用四台超声多普勒仪(Acuson、ATL、日立阿洛卡、东芝)对15例患者独立测量超声多普勒参数。15例患者中的8例在15天后由相同操作人员仅使用一台仪器重新检查。获得了PBV、PBF和CI的显著不同值。PBV的变异性与设备相关(32%),与操作人员相关的比例较小(5%)。未发现与检查时间相关的系统性影响。在培训以确定精确方案后,四名操作人员对8例不同患者进行了新的测量。各操作人员之间的任何参数均未发现显著差异,PBV的95%置信区间(CL)和变异系数(CV)显著降低(CL从±26.4%降至±15.6%)。这些结果表明:(1)不同设备的多普勒测量之间存在显著的系统性变异性;(2)多普勒测量不存在显著的随时间的系统性变异性;(3)合作培训计划可降低直接测量(如PBV)的观察者间变异性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验