Switching from DXA pencil-beam to fan-beam. I: Studies in vitro at four centers.

作者信息

Eiken P, Bärenholdt O, Bjørn Jensen L, Gram J, Pors Nielsen S

机构信息

Department of Clinical Physiology, Hillerød Hospital, Denmark.

出版信息

Bone. 1994 Nov-Dec;15(6):667-70. doi: 10.1016/8756-3282(94)90315-8.

Abstract

The performance of the Hologic QDR-2000 DXA osteodensitometer was critically evaluated at four centers, using at all four centers one bone equivalent humanoid spine phantom supplied by the manufacturer. Results were compared with results from Hologic QDR-1000/W using that phantom tested at the same centers. It appears that the concept of fan-beam scanning--as used in the QDR-2000: a fan-beam, a linear array detector above the phantom, and an x-ray tube located rather close to the spine below the phantom--creates problems due to the magnification effect of the fan beam. The effect of decreasing the distance between the "vertebrae" of the phantom and the couch are: bone mineral content (BMC) increases by 2.8% per cm, projected area (Area) by 2.8% per cm, and bone mineral density (BMD) is unchanged. When QDR-1000/W is upgraded to QDR-2000, BMD is relatively constant, but there are shifts of BMC and Area which are partly due to the magnification effect of the fan-beam. Replacement of a QDR-1000/W with a QDR-2000 can invalidate longitudinal measurements, even for BMD, unless the proportionality factors of the QDR-2000 are checked and, if necessary, changed. This is true for switching from QDR-1000/W to pencil-beam mode of QDR-2000 or to fan-beam mode of QDR-2000. Even with pencil-beam mode, the long-term precision error with phantoms is higher for QDR-2000 than for QDR-1000/W (for BMD, 0.47% versus 0.35%).

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验