Evaluation of reproducibility of a vessel wall movement detector system for assessment of large artery properties.

作者信息

Kool M J, van Merode T, Reneman R S, Hoeks A P, Struyker Boudier H A, Van Bortel L M

机构信息

Cardiovascular Research Institute, University of Limburg, Maastricht, The Netherlands.

出版信息

Cardiovasc Res. 1994 May;28(5):610-4. doi: 10.1093/cvr/28.5.610.

Abstract

OBJECTIVE

Arterial distensibility (DC) and compliance (CC) are vessel wall properties of large arteries that can be measured non-invasively with a custom made vessel wall movement detector system (VWMDS). This study investigated the reproducibility of this device in 10 volunteers.

METHODS

To investigate intraobserver intrasession and intraobserver intersession variability, arterial diameter (D) and relative change in diameter during the heart cycle (delta D/D) were measured in the elastic common carotid artery, and in the muscular femoral and brachial arteries. Interobserver intrasession variability was examined in common carotid artery by two observers, while interobserver variability on the same image was assessed for common carotid and femoral arteries. Variability was expressed as the coefficient of variation.

RESULTS

For common carotid artery, intraobserver intrasession variability was 7.9(SEM 1.6)% (delta D/D), 4.5(1.1)% (D), 8.3(1.3)% (DC), and 9.1(2.6)% (CC), respectively. In femoral artery it was 12.4(2.2)% (delta D/D), 2.7(0.6)% (DC), 13.4(2.2)% (DC), and 12.5(2.1)% (CC). For brachial artery it was 13.4(2.8)% (delta D/D), 2.5(0.5)% (D), 16.1(2.5)% (DC), and 15.6(2.6)% (CC). Intraobserver intersession variability was comparable to intraobserver intrasession variability for all vessels. Interobserver intrasession variability for common carotid artery was 11.3(2.6)% (delta D/D) and 8.6(1.9)% (D), but was larger for DC and CC. Interobserver variability on the same image was < 5% for common carotid and femoral arteries.

CONCLUSIONS

In conclusion, the vessel wall movement detector system has a good technical reproducibility. Intraobserver intrasession and intersession variability are comparable, and are larger in muscular arteries. This might be due to a larger variation in tone of these arteries, which are under permanent neurohumoral control. Interobserver intrasession variability was larger than intraobserver variability and might be influenced by differences in observers' skill and spontaneous variation in vessel wall properties.

摘要

目的

动脉扩张性(DC)和顺应性(CC)是大动脉的血管壁特性,可使用定制的血管壁运动检测系统(VWMDS)进行无创测量。本研究调查了该设备在10名志愿者中的可重复性。

方法

为研究观察者内单次测量和观察者内不同测量之间的变异性,在弹性的颈总动脉、肌性的股动脉和肱动脉中测量动脉直径(D)和心动周期中直径的相对变化(δD/D)。两名观察者在颈总动脉中检测观察者间单次测量的变异性,同时评估颈总动脉和股动脉在同一图像上的观察者间变异性。变异性用变异系数表示。

结果

对于颈总动脉,观察者内单次测量的变异性分别为7.9(标准误1.6)%(δD/D)、4.5(1.1)%(D)、8.3(1.3)%(DC)和9.1(2.6)%(CC)。在股动脉中,分别为12.4(2.2)%(δD/D)、2.7(0.6)%(D)、13.4(2.2)%(DC)和12.5(2.1)%(CC)。在肱动脉中,分别为13.4(2.8)%(δD/D)、2.5(0.5)%(D)、16.1(2.5)%(DC)和15.6(2.6)%(CC)。所有血管的观察者内不同测量之间的变异性与观察者内单次测量的变异性相当。颈总动脉的观察者间单次测量变异性为11.3(2.6)%(δD/D)和8.6(1.9)%(D),但DC和CC的变异性更大。颈总动脉和股动脉在同一图像上的观察者间变异性<5%。

结论

总之,血管壁运动检测系统具有良好的技术可重复性。观察者内单次测量和不同测量之间的变异性相当,且在肌性动脉中更大。这可能是由于这些动脉的张力变化更大,它们处于持续的神经体液控制之下。观察者间单次测量的变异性大于观察者内变异性,可能受观察者技能差异和血管壁特性的自发变化影响。