J Biomed Mater Res. 1994 May;28(5):563-72. doi: 10.1002/jbm.820280506.

Abstract

A tensile test has been developed to test bioactive coating materials. Hydroxylapatite (HA) and fluorapatite (FA) coatings with various roughnesses were tested using sandblasted titanium (Ti) as a control material. Twelve goats received 7 implants each for 6 (2 goats), 12 (5), and 24 (4) week implantation periods. After 12 weeks the mean tensile strength values were highest for polished hydroxylapatite (HAP) followed by HA > FAP (polished fluorapatite) > FA > Ti, and hydroxylapatite type coatings were found to exhibit significantly higher values than fluorapatite type coatings or titanium implants. After 24 weeks no statistically significant differences could be found between any of the implant types. The order of the mean strength values was now HAP > HA > FA > FAP > Ti. Fracture always occurred between the coating and titanium if bone contact had been established. In conclusion, it is suggested that results from different types of test methods cannot be used to compare different types of bioactive coatings.

摘要

已开发出一种拉伸试验来测试生物活性涂层材料。使用喷砂处理的钛(Ti)作为对照材料,对具有不同粗糙度的羟基磷灰石(HA)和氟磷灰石(FA)涂层进行了测试。12只山羊,每只植入7枚植入物,植入期分别为6周(2只山羊)、12周(5只)和24周(4只)。12周后,抛光羟基磷灰石(HAP)的平均拉伸强度值最高,其次是HA>FAP(抛光氟磷灰石)>FA>Ti,并且发现羟基磷灰石类涂层的值显著高于氟磷灰石类涂层或钛植入物。24周后,在任何植入物类型之间均未发现统计学上的显著差异。现在平均强度值的顺序为HAP>HA>FA>FAP>Ti。如果已建立骨接触,则断裂总是发生在涂层与钛之间。总之,建议不能使用不同类型测试方法的结果来比较不同类型的生物活性涂层。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验