An electronic method for rapid measurement of haematocrit in blood samples.

作者信息

Cha K, Faris R G, Brown E F, Wilmore D W

机构信息

Department of Surgery, Brigham and Women's Hospital, Boston, MA 02115.

出版信息

Physiol Meas. 1994 May;15(2):129-37. doi: 10.1088/0967-3334/15/2/003.

Abstract

The bioelectrical impedance method was used to determine the haematocrit in patient blood samples. The resistance (R) and reactance (X) of a constant volume of blood were measured at a low (50 kHz) and high (1 MHz) frequency. These dual-frequency impedance measurements were then used to determine the intracellular and extracellular (plasma) fluid volume of the blood and thus the haematocrit in 308 blood samples. The values compared favourably with the microhaematocrit determined on the same samples, giving a correlation coefficient of 0.989 and a standard error estimation of 1.41%. Reactance at 1 MHz was a better parameter for predicting haematocrit than resistances. Factors affecting the resistivity of plasma, such as electrolyte and protein concentrations, were the error sources in the impedance technique. Our technique reduced such errors significantly. In conclusion, this electronic method can be used to measure haematocrit accurately and may be more desirable than conventional methods in certain clinical situations.