Suppr超能文献

陶瓷流量传感器的体温、气压、饱和水蒸气压校正

BTPS correction for ceramic flow sensor.

作者信息

Hankinson J L, Viola J O, Petsonk E L, Ebeling T R

机构信息

Centers for Disease Control, National Institute for Occupational Safety and Health-ALOSH, Morgantown, WVa 26505.

出版信息

Chest. 1994 May;105(5):1481-6. doi: 10.1378/chest.105.5.1481.

Abstract

Several commercially available spirometers use unheated ceramic elements as flow sensors to determine flow and calculate volume of air. The usual method of correcting the resulting flow and volume values to body temperature pressure saturated (BTPS) is to apply a constant factor approximately equal to 30 percent of the full BTPS correction factor. To evaluate the usual BTPS correction factor technique, we tested several sensors with a mechanical pump using both room air and air heated to 37 degrees C and saturated with water vapor. The volume signals used to test the sensors were volume ramps (constant flow) and the first four American Thoracic Society (ATS) standard waveforms. The percent difference in FEV1 obtained using room vs heated-humidified air (proportional to the magnitude of the BTPS correction factor needed) ranged from 0.3 percent to 6.2 percent and varied with the number of maneuvers previously performed, the time interval between maneuvers, the volume of the current and previous maneuvers, and the starting temperature of the sensor. The temperature of the air leaving the sensor (exit temperature) showed a steady rise with each successive maneuver using heated air. When six subjects performed repeated tests over several days (each test consisting of at least three maneuvers), a maneuver order effect was observed similar to the results using the mechanical pump. These results suggest that a dynamic, rather than static, BTPS correction factor is needed for accurate estimations of forced expiratory volumes and to reduce erroneous variability between successive maneuvers. Use of exit air temperature provides a means of estimating a dynamic BTPS correction factor, and this technique may be sufficient to provide an FEV1 accuracy of less than +/- 3 percent for exit air temperatures from 5 degrees to 28 degrees C.

摘要

几种市面上可买到的肺活量计使用未加热的陶瓷元件作为流量传感器来测定流量并计算空气体积。将所得流量和体积值校正到体温、气压饱和(BTPS)状态的常用方法是应用一个约等于完整BTPS校正因子30%的常数因子。为了评估常用的BTPS校正因子技术,我们使用机械泵,分别用室温和加热到37摄氏度并饱和水蒸气的空气对几种传感器进行了测试。用于测试传感器的体积信号是体积斜坡(恒定流量)和美国胸科学会(ATS)的前四种标准波形。使用室温空气与加热加湿空气(与所需BTPS校正因子的大小成比例)获得的第一秒用力呼气容积(FEV1)的百分比差异在0.3%至6.2%之间,并且随之前进行的操作次数、操作之间的时间间隔、当前和之前操作的体积以及传感器的起始温度而变化。使用加热空气时,离开传感器的空气温度(出口温度)在每次连续操作时都呈现稳定上升。当六名受试者在几天内进行重复测试(每次测试至少包括三次操作)时,观察到一种操作顺序效应,类似于使用机械泵时的结果。这些结果表明,为了准确估计用力呼气容积并减少连续操作之间的错误变异性,需要一个动态而非静态的BTPS校正因子。使用出口空气温度提供了一种估计动态BTPS校正因子的方法,并且这种技术可能足以在出口空气温度为5摄氏度至28摄氏度时提供小于+/- 3%的FEV1准确度。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验