A bovine model was developed for biomechanical evaluation of anterior cruciate ligament (ACL) reconstruction using patellar bone-tendon-bone (b-t-b) autograft to examine the differences in time zero fixation mechanical properties of different interference screw lengths and diameters. The surgical technique of interference screw fixation of the b-t-b complex performed clinically was reproduced in a controlled animal model. The femur-patellar tendon graft-tibia complex was tested with anterior displacement of the tibia in 30 degrees of knee flexion to allow examination of the femoral and tibial fixation properties simultaneously. The statistical model concurrently explored differences between screw length and diameter while accounting for variations between graft properties. No statistically significant differences were found between the 7- and 9-mm screws with respect to peak load or energy to failure when using a 10-mm triangular graft in a 10-mm tunnel. The 7- and 9-mm screws were superior to the 5.5-mm screws with respect to these same parameters. Based on our results, the 7-mm interference screws can be used with equal confidence as the 9-mm screw, and the 20-mm length can be similarly exchanged for 30-mm length for patellar b-t-b graft fixation.
建立了一种牛模型,用于对使用髌腱骨(b-t-b)自体移植物进行前交叉韧带(ACL)重建的生物力学评估,以检查不同长度和直径的干涉螺钉在初始固定时力学性能的差异。在一个可控的动物模型中重现了临床上对b-t-b复合体进行干涉螺钉固定的手术技术。在膝关节屈曲30度时,对股骨-髌腱移植物-胫骨复合体进行胫骨前移测试,以便同时检查股骨和胫骨的固定性能。统计模型在考虑移植物性能差异的同时,还探讨了螺钉长度和直径之间的差异。当在10毫米隧道中使用10毫米三角形移植物时,7毫米和9毫米螺钉在峰值负荷或破坏能量方面没有统计学上的显著差异。在这些相同参数方面,7毫米和9毫米螺钉优于5.5毫米螺钉。根据我们的结果,7毫米干涉螺钉可以与9毫米螺钉一样放心使用,对于髌腱骨移植物固定,20毫米的长度可以同样换成30毫米的长度。