Suppr超能文献

用于心脏导管插入实验室的两种血氧饱和度测定仪的评估。

Evaluation of two oximeters for use in cardiac catheterization laboratories.

作者信息

Freeman G L, Steinke J M

机构信息

Department of Medicine, University of Texas Health Science Center, San Antonio 78284-7756.

出版信息

Cathet Cardiovasc Diagn. 1993 Sep;30(1):51-7. doi: 10.1002/ccd.1810300114.

Abstract

We evaluated two whole-blood oximeters designed for use in the cardiac catheterization laboratory: the Oxicom 3000 and the AVOXimeter 1000. Unlike the larger CO-Oximeters, which hemolyze a blood sample before analysis, these simple instruments use disposable cuvettes to allow determination of oxyhemoglobin saturation of whole blood. Thus they eliminate the need for cleaning solutions and considerable maintenance. We evaluated the accuracy of these instruments in comparison with a Radiometer OSM3 and found them both to work well on routine samples. Of note, the AVOXimeter performed better than the Oxicom when the sample was hemolyzed and when green dye was present. In addition, the AVOXimeter measures the total hemoglobin concentration and the oxygen content of each sample, eliminating errors that hemodilution introduces into the determination of cardiac output by the Fick principle. We conclude that whole-blood oximeters are accurate and useful instruments for use in the cardiac catheterization laboratory.

摘要

我们评估了两款设计用于心导管实验室的全血血氧计

Oxicom 3000和AVOXimeter 1000。与在分析前会使血样溶血的较大型的联合血氧计不同,这些简易仪器使用一次性比色皿来测定全血的氧合血红蛋白饱和度。因此,它们无需清洁溶液,维护量也大大减少。我们将这些仪器的准确性与Radiometer OSM3进行了比较,发现它们在常规样本上的表现都很好。值得注意的是,当样本溶血或存在绿色染料时,AVOXimeter的表现优于Oxicom。此外,AVOXimeter可测量每个样本的总血红蛋白浓度和氧含量,消除了血液稀释通过菲克原理测定心输出量时引入的误差。我们得出结论,全血血氧计是用于心导管实验室的准确且实用的仪器。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验