Arditi A, Cagenello R
Vision Research Laboratory, Lighthouse Inc., New York, New York 10017.
Invest Ophthalmol Vis Sci. 1993 Jan;34(1):120-9.
To determine boundary values of test-retest reliability and sensitivity to acuity change that are unlikely to be exceeded in any clinical situation, for a popular visual acuity chart (Lighthouse/ETDRS) with three scoring methods; to discuss general methodological issues associated with statistical accuracy of optotype chart testing; and to link measures of test reliability to measures of sensitivity to change.
Five highly practiced subjects were tested using a computer-controlled acuity testing system in a procedure designed to reduce measurement error. Subjects read the computerized chart 156 times, yielding a sample of 78 test-retest comparisons.
Under conditions likely to minimize variability, visual acuity may, with 95% confidence, be ascertained only within +/- 0.1 log units, using this chart with the recommended letter-by-letter scoring. Detecting a significant change in visual acuity requires about +/- 0.14 log units for the same degree of confidence.
These measurements may be viewed as approaching the upper limit of reliability of this letter chart. Reliability probably is considerably less in typical usage.