Onks D, Silverman L, Robertson A
Department of Pediatrics, East Carolina University, Greenville, North Carolina.
Acta Paediatr. 1993 Jan;82(1):19-21. doi: 10.1111/j.1651-2227.1993.tb12507.x.
To determine the effect of skin pigments on transcutaneous bilirubinometer readings, we measured the effect of bilirubin, melanin, and oxyhemoglobin solutions on transcutaneous bilirubinometer readings in vitro. Our results showed that the variability of the readings in vitro was related to the instrument's non-linear response to bilirubin and melanin concentration and an inaccurate oxyhemoglobin correction factor. These factors should be considered in developing a more accurate non-invasive method of monitoring serum bilirubin concentration.
为了确定皮肤色素对经皮胆红素仪读数的影响,我们在体外测量了胆红素、黑色素和氧合血红蛋白溶液对经皮胆红素仪读数的影响。我们的结果表明,体外读数的变异性与仪器对胆红素和黑色素浓度的非线性反应以及不准确的氧合血红蛋白校正因子有关。在开发一种更准确的无创监测血清胆红素浓度的方法时,应考虑这些因素。