Comparative study of power-based versus mean frequency-based transcranial color-coded duplex sonography in normal adults.

作者信息

Baumgartner R W, Schmid C, Baumgartner I

机构信息

Department of Neurology, University of Bern, Inselspital, Switzerland.

出版信息

Stroke. 1996 Jan;27(1):101-4. doi: 10.1161/01.str.27.1.101.

Abstract

BACKGROUND AND PURPOSE

Power-based transcranial color-coded duplex sonography (p-TCCD) is a new ultrasonic method that has advantages compared with frequency-based TCCD (f-TCCD), since it is essentially independent of the angle of insonation, not subject to aliasing, and has a better signal-to-noise ratio. The purpose of this study was to evaluate the ability of p-TCCD to visualize flow in cerebral parenchyma and to compare the advantages, limitations, and reliability of velocity measurements of p-TCCD versus f-TCCD in the major basal cerebral arteries of normal subjects.

METHODS

Two investigators performed 15 p-TCCD and 15 f-TCCD studies in 30 normal subjects with adequate ultrasonic windows. Each investigator did a p-TCCD or f-TCCD study in every patient, and each was blinded in every case to the results of the other. Peak systolic (Vs) and end-diastolic (Vd) velocities were determined in the anterior, middle, and posterior cerebral, basilar, and vertebral arteries. The reliability of p-TCCD velocimetry was evaluated by calculating both the correlation coefficient (r) of the difference between p-TCCD versus f-TCCD measurements and the coefficient of variation (CV), defined as the difference between the mean values for p-TCCD and f-TCCD divided by the mean values for f-TCCD measurements, expressed as percent.

RESULTS

p-TCCD did not display flow in cerebral parenchyma but depicted arteries with a course perpendicular to that of the ultrasound beam and showed good reliability of velocity measurements in all examined arteries: for Vs, r was .84 to .93 (P < .001) and CV was 7.7% to 10.8%; for Vd, r was .87 to .90 (P < .001) and CV was 10.3% to 13.7%. The lack of directional and velocity information and tissue motion artifacts were unimportant limitations in p-TCCD.

CONCLUSIONS

Compared with f-TCCD, p-TCCD had no important advantages but had several unimportant limitations in a study of normal adults with adequate ultrasonic windows.

摘要

背景与目的

基于功率的经颅彩色编码双功能超声(p-TCCD)是一种新型超声检查方法,与基于频率的经颅彩色编码双功能超声(f-TCCD)相比具有优势,因为它基本不受声束入射角度的影响,不易出现混叠现象,且具有更好的信噪比。本研究旨在评估p-TCCD显示脑实质内血流的能力,并比较p-TCCD与f-TCCD在正常受试者大脑主要基底动脉中进行速度测量的优势、局限性及可靠性。

方法

两名研究人员对30例具有足够超声窗的正常受试者进行了15次p-TCCD和15次f-TCCD检查。每位研究人员对每位患者进行一次p-TCCD或f-TCCD检查,并且在每种情况下对另一位研究人员的检查结果均不知情。测定大脑前动脉、大脑中动脉、大脑后动脉、基底动脉和椎动脉的收缩期峰值速度(Vs)和舒张末期速度(Vd)。通过计算p-TCCD与f-TCCD测量值之间差异的相关系数(r)以及变异系数(CV)来评估p-TCCD测速的可靠性,CV定义为p-TCCD与f-TCCD平均值之差除以f-TCCD测量值的平均值,以百分比表示。

结果

p-TCCD未显示脑实质内的血流,但描绘了与超声束垂直走行的动脉,并且在所有检查的动脉中均显示出良好的速度测量可靠性:对于Vs,r为0.84至0.93(P < 0.001),CV为7.7%至10.8%;对于Vd,r为0.87至0.90(P < 0.001),CV为10.3%至13.7%。缺乏方向和速度信息以及组织运动伪像在p-TCCD中是不太重要的局限性。

结论

在具有足够超声窗的正常成年人研究中,与f-TCCD相比,p-TCCD没有重要优势,但有一些不太重要的局限性。