带螺钉固定的多孔涂层髋臼组件。5至10年的结果。

Porous-coated acetabular components with screw fixation. Five to ten-year results.

作者信息

Latimer H A, Lachiewicz P F

机构信息

Department of Orthopaedics, University of North Carolina at Chapel Hill, USA.

出版信息

J Bone Joint Surg Am. 1996 Jul;78(7):975-81. doi: 10.2106/00004623-199607000-00001.

Abstract

The results of 136 consecutive primary total hip arthroplasties performed by one surgeon with the Harris-Galante-I porous-coated acetabular component were reviewed at a mean of seven years (range, five to ten years). In all hips, the outer diameter of the acetabular component was the same as the diameter of the final reamer used in the preparation of the acetabulum. However, this reamer was used only briefly at the rim of the acetabulum, and therefore the components had so-called press-fit stability. A mean of four screws (range, three to six screws) were used for additional fixation of the component. The clinical evaluation was performed with use of the Harris hip score. Standardized anteroposterior radiographs of the pelvis were assessed for migration of the component, radiolucent and radiodense lines, linear wear of the polyethylene, and osteolysis. No acetabular component had been revised for loosening and none were radiographically loose at the time of the most recent follow-up evaluation. There were no complications related to the use of the screws, and no screw had bent or broken. A non-progressive radiolucent line was seen in one acetabular zone in thirty-four hips (25 per cent) and in two acetabular zones in six hips (4 per cent). No hip had a radiolucent line in all three acetabular zones. The mean rate of linear wear of the polyethylene was 0.1 millimeter per year. There was no dissociation of the acetabular liner from the metal shell. Two hips (1 per cent) had asymptomatic osteolysis in the ischium and adjacent to the rim of the acetabular component; this was treated with grafting at the site of the lesion and exchange of the femoral head and the worn polyethylene liner. Five femoral components inserted without cement and one inserted with cement were revised because of loosening. The data suggest that, at a mean of seven years, fixation of this porous-coated component was uniformly excellent. The low prevalence of radiolucent lines and the absence of loosening are probably related, in part, to the technique of implantation. The low prevalence of polyethylene wear and pelvic osteolysis is a notable improvement compared with the results of arthroplasty with other porous-coated acetabular components. Although the results of the present study are encouraging, longer follow-up is necessary to determine the prevalence of late loosening and osteolysis.

摘要

对一位外科医生使用Harris-Galante-I多孔涂层髋臼组件连续进行的136例初次全髋关节置换术的结果进行了回顾,平均随访时间为7年(范围为5至10年)。在所有髋关节中,髋臼组件外径与髋臼准备中使用的最终扩孔钻直径相同。然而,该扩孔钻仅在髋臼边缘短暂使用,因此组件具有所谓的压配稳定性。平均使用4枚螺钉(范围为3至6枚螺钉)对组件进行额外固定。采用Harris髋关节评分进行临床评估。对骨盆的标准前后位X线片进行评估,以观察组件的移位、透光线和致密线、聚乙烯的线性磨损以及骨溶解情况。在最近一次随访评估时,没有髋臼组件因松动而翻修,且影像学检查均未发现松动。未出现与螺钉使用相关的并发症,也没有螺钉弯曲或折断。在34例髋关节(25%)的一个髋臼区域以及6例髋关节(4%)的两个髋臼区域可见一条非进行性透光线。没有髋关节在所有三个髋臼区域均出现透光线。聚乙烯的平均线性磨损率为每年0.1毫米。髋臼内衬与金属外壳未分离。2例髋关节(1%)在坐骨及髋臼组件边缘附近出现无症状性骨溶解;在病变部位进行植骨,并更换股骨头和磨损的聚乙烯内衬进行治疗。5例未使用骨水泥植入的股骨组件和1例使用骨水泥植入的股骨组件因松动而翻修。数据表明,平均7年时,这种多孔涂层组件的固定效果总体优异。透光线的低发生率和无松动情况可能部分与植入技术有关。与其他多孔涂层髋臼组件的关节置换结果相比,聚乙烯磨损和骨盆骨溶解的低发生率是一个显著改善。尽管本研究结果令人鼓舞,但仍需要更长时间的随访来确定晚期松动和骨溶解的发生率。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验