Suppr超能文献

使用改进的六氟化硫(SF6)色谱分析法测量心输出量。

Measurement of cardiac output using improved chromatographic analysis of sulfur hexafluoride (SF6).

作者信息

Klocke F J, Roberts D L, Farhi E R, Naughton B J, Sekovski B, Klocke R A

出版信息

Respir Physiol. 1977 Jun;30(1-2):99-107. doi: 10.1016/0034-5687(77)90024-x.

Abstract

A constant current variable frequency pulsed electron capture detector has been incorporated into the gas chromatographic analysis of trace amounts of sulfur hexafluoride (SF6) in water and blood. The resulting system offers a broader effective operating range than more conventional electron capture units and has been utilized for measurements of cardiac output employing constant-rate infusion of dissolved SF6. The SF6 technique has been validated against direct volumetric measurements of cardiac output in a canine right-heart bypass preparation and used subsequently for rapidly repeated measurements in conscious animals and man.

摘要

一种恒流变频脉冲电子捕获检测器已被应用于气相色谱分析水中和血液中的痕量六氟化硫(SF6)。与更传统的电子捕获装置相比,该系统具有更宽的有效工作范围,并已用于通过恒速输注溶解的SF6来测量心输出量。在犬右心旁路制备中,已针对心输出量的直接体积测量对SF6技术进行了验证,随后用于清醒动物和人类的快速重复测量。