外周定量计算机断层扫描的体外准确性和精密度研究。

Accuracy and precision study in vitro for peripheral quantitative computed tomography.

作者信息

Takada M, Engelke K, Hagiwara S, Grampp S, Genant H K

机构信息

Osteoporosis Research Group, University of California, San Francisco 94143-0628, USA.

出版信息

Osteoporos Int. 1996;6(3):207-12. doi: 10.1007/BF01622736.

Abstract

We evaluated the accuracy and precision of a peripheral quantitative computed tomography (pQCT) scanner, the Stratec XCT-960, using 12 human cadaveric forearms. The accuracy was determined by comparing the total bone mineral content (BMC) with the ash weight (AW). We scanned and ashed three consecutive slices (thickness 2.5 mm) at the standard position (s-position) and at 2.5 mm both proximal and distal to the s-position. The correlation coefficient between the AW and total BMC using slices at the s-position was r = 0.87 with an accuracy error (random component) of 15.5%. The correlation coefficient using all slices was r = 0.90 with an accuracy error of 14.3%. The correlation coefficient improved to r = 0.95 with an accuracy error of 9.7% after averaging the results of all three slices for each forearm. The short-term precision error expressed as the coefficient of variation (CV) of bone mineral density (BMD) and BMC was determined by measuring the forearms five times either with repositioning or without repositioning. The CVs with repositioning were 2.77 and 1.15 for total BMD and BMC, 1.85 for trabecular BMD; without repositioning they were 0.29, 0.58 and 0.69 respectively. To further evaluate the influence of positioning, additional scans were performed at 1, 2 and 5 mm proximal, and 1 and 2 mm distal to the s-position. BMD and BMC were greatly influenced by the scan location; for example, the percentage differences in trabecular BMD 1 mm distal and proximal relative to the s-position were 2.5% +/- 5.1% and 0.18% +/- 6.3%, respectively. The Stratec XCT-960 appears to be a moderately accurate and highly precise scanner with potential usefulness for evaluating BMC and BMD of ultradistal radius.

摘要

我们使用12具人类尸体前臂评估了外周定量计算机断层扫描(pQCT)扫描仪Stratec XCT - 960的准确性和精密度。通过将总骨矿物质含量(BMC)与灰重(AW)进行比较来确定准确性。我们在标准位置(s位置)以及在s位置近端和远端2.5毫米处扫描并灰化了三个连续的切片(厚度2.5毫米)。使用s位置的切片时,AW与总BMC之间的相关系数为r = 0.87,准确性误差(随机成分)为15.5%。使用所有切片时相关系数为r = 0.90,准确性误差为14.3%。在对每个前臂的所有三个切片结果进行平均后,相关系数提高到r = 0.95,准确性误差为9.7%。短期精密度误差以骨矿物质密度(BMD)和BMC的变异系数(CV)表示,通过对前臂进行五次测量来确定,测量时要么重新定位要么不重新定位。重新定位时,总BMD和BMC的CV分别为2.77和1.15,小梁BMD的CV为1.85;不重新定位时,它们分别为0.29、0.58和0.69。为了进一步评估定位的影响,在s位置近端1、2和5毫米以及远端1和2毫米处进行了额外扫描。BMD和BMC受扫描位置的影响很大;例如,相对于s位置,小梁BMD在远端和近端1毫米处的百分比差异分别为2.5%±5.1%和0.18%±6.3%。Stratec XCT - 960似乎是一台准确性中等且精密度高的扫描仪,在评估桡骨远端的BMC和BMD方面具有潜在用途。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验