Verdun F R, Meuli R A, Bochud F O, Imsand C, Raimondi S, Schnyder P, Valley J F
Institut de Radiophysique Appliquée, Centre Universitaire, Lausanne, Switzerland.
Eur Radiol. 1996;6(4):485-8. doi: 10.1007/BF00182477.
Image quality and dose produced by a spiral CT with various pitch values were investigated on the basis of test-object images and measurements of CT dose indexes. No major difference in image quality or dose was observed when comparing sequential and spiral mode acquisition with a pitch of one. Increase in pitch value produces a loss of contrast which leads to a loss of detectability. Nevertheless, in realistic protocols the image quality in the transverse plane remains acceptable up to pitch values of 1.3-1.6.
基于测试物体图像和CT剂量指数测量,研究了不同螺距值的螺旋CT产生的图像质量和剂量。当比较螺距为1的序列扫描模式和螺旋扫描模式时,未观察到图像质量或剂量有重大差异。螺距值增加会导致对比度降低,进而导致可检测性下降。然而,在实际扫描方案中,直到螺距值达到1.3 - 1.6,横断面的图像质量仍可接受。