Cohnen M, Fischer H, Hamacher J, Lins E, Kötter R, Mödder U
Institute of Diagnostic Radiology, Heinrich-Heine University, Düsseldorf, Germany.
AJNR Am J Neuroradiol. 2000 Oct;21(9):1654-60.
CT is a frequent examination that is performed using ionizing radiation. We sought to assess image-quality changes on CT scans of the head when the radiation dose is reduced by changing tube current and kilovoltage.
A formalin-fixed cadaver was examined in conventional and helical mode by use of two CT-scanners. Surface dose was measured with standard scanning parameters, and after reduction of tube current and kilovoltage. Five experienced examiners independently evaluated subjective image quality.
In the conventional mode, the highest surface dose was 83.2 mGy (scanner 1: helical mode, 55.6 mGy), and 66.0 mGy (scanner 2: helical mode, 55.9 mGy). By changing kVp and mAs, a dose reduction of up to 75% (scanner 1), and 60% (scanner 2) was achieved. No observable differences in image quality between scans obtained with doses from 100% to 60% of standard settings were noted. Ten of 20 images obtained with the highest dose and 13 of 20 images obtained with lowest dose (19-29.4 mGy) were reliably identified by subjective quality assessment. Scans produced with a surface dose of less than 30 mGy were judged uninterpretable.
Standard parameters used in cranial CT are oriented toward best image quality. A dose reduction up to 40% may be possible without loss of diagnostic image quality.
CT是一种常用的使用电离辐射的检查方法。我们试图评估通过改变管电流和千伏电压来降低辐射剂量时,头部CT扫描图像质量的变化。
使用两台CT扫描仪以传统模式和螺旋模式对一具福尔马林固定的尸体进行检查。用标准扫描参数测量表面剂量,并在降低管电流和千伏电压后测量。五名经验丰富的检查人员独立评估主观图像质量。
在传统模式下,最高表面剂量为83.2 mGy(扫描仪1:螺旋模式,55.6 mGy),以及66.0 mGy(扫描仪2:螺旋模式,55.9 mGy)。通过改变千伏峰值和毫安秒,剂量分别降低了高达75%(扫描仪1)和60%(扫描仪2)。在标准设置剂量的100%至60%之间获得的扫描图像质量上未观察到明显差异。通过主观质量评估,在最高剂量获得的20张图像中有10张以及在最低剂量(19 - 29.4 mGy)获得的20张图像中有13张被可靠识别。表面剂量小于30 mGy的扫描被判定为无法解读。
头颅CT使用的标准参数旨在实现最佳图像质量。在不损失诊断图像质量的情况下,剂量可能降低高达40%。